Use raw connectivity, such as coherence, in protocols and live neurofeedback without a norm database. A Connectivity bar in a protocol shows the connectivity between all pairs of its selected positions in the chosen frequency band.
During training the value is estimated over 2-second windows of the last 16 seconds; in calibration, Studio, reports and multi-session graphs over the whole segment. Formulas and report scores offer the same calculation types.
The protocol creator has a new Synchrony (sum channel) bar: it averages the signals of the selected positions and trains the amplitude of that average in the chosen band, so the bar only rises when all positions are active at the same time and in phase, as in Open Focus training.
Formulas and report scores offer the same calculation as Sum channel amplitude over a selection of positions. It uses the reference of the session (hardware reference or live re-referencing, in Studio the selected montage), and the feature info of the bar explains the calculation.
The protocol creator has a new Connectivity bar, and formulas and report scores offer Coherence, PLI, wPLI and PLV as calculation types over a selection of positions. The value is the connectivity between all pairs of the selected positions in the chosen frequency band, estimated over 2-second windows of the last 16 seconds during training and over the whole segment in calibration, Studio, reports and multi-session graphs.
Connectivity bars show one title such as “Alpha (Coherence)”; a custom title replaces it completely.
Live neurofeedback now uses the newest EEG sample immediately instead of one sample late.